Key result
Transvenous lead extraction using the TightRail Sub-C extraction sheath achieved complete procedural success in 95.6% and clinical procedural success in 97.8% of patients.
Why the study?
To evaluate the efficacy and safety of the novel short rotating dilator sheath TightRail Sub-C in transvenous lead extraction.
Is transvenous lead extraction using the TightRail TM Sub-C extraction sheath safe and effective in patients requiring device lead management?
Observational (n=45)
No
Is transvenous lead extraction using the TightRail TM Sub-C extraction sheath safe and effective in patients requiring device lead management?
The routine use of the TightRail TM Sub-C extraction sheath for transvenous lead extraction is associated with high procedural success rates and an acceptable safety profile.
May support TightRail Sub-C consideration in lead extraction; leaves open need for prospective comparative trials.
Introduction Transvenous lead extraction has become a frequent procedure in the context of device lead management and various tools are available. The aim of this study was to investigate the efficacy and safety of the novel short rotating dilator sheath TightRail TM Sub‐C (Sub‐C) in transvenous lead extraction. Methods For this retrospective single‐center analysis, we included consecutive patients undergoing transvenous lead extraction using the Sub‐C from January 2018 until February 2020 at the University Heart Center Zurich. Results A total of 87 leads were extracted within 45 patients using the Sub‐C extraction sheath. The mean dwell time of the leads was 112 ± 9.1 months. Complete procedural success was achieved in 95.6% (43/45) and clinical procedural success in 97.8% (44/45). Two major complications (4.4%, 2/45) occurred, but neither was directly related to the Sub‐C. Conclusion This retrospective single‐center analysis suggests that transvenous lead extraction with a routine use of the TightRail TM Sub‐C extraction sheath is a safe strategy resulting in high success rates that may offer useful theoretical benefits. Future studies are needed to evaluate the incremental benefit of a routine use of short extraction sheaths including the Sub‐C during TLE procedures.
No takes yet. Share an insight, caveat, or question.
Frei et al. (2023) conducted an observational in Transvenous lead extraction (n=45). TightRail Sub-C extraction sheath was evaluated on Complete procedural success. Transvenous lead extraction using the TightRail Sub-C extraction sheath achieved complete procedural success in 95.6% and clinical procedural success in 97.8% of patients.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: