Key result
Newer mechanical extraction equipment (Evolution and Evolution RL) progressively increased the rate of complete lead extraction from 88.0% to 97.6%.
Why the study?
Does the use of newer mechanical extraction equipment improve complete extraction rates in patients undergoing transvenous lead extraction compared to older equipment?
Cohort (n=288)
Does the use of newer mechanical extraction equipment improve complete extraction rates in patients undergoing transvenous lead extraction compared to older equipment?
Newer mechanical extraction tools (Evolution and Evolution RL) significantly improve the complete extraction rate of older and more complex pacemaker/ICD leads with a good safety profile.
Answers are grounded in this paper and cite their sources.
Supports the use of newer mechanical sheaths for complex ICD and pacemaker lead removals;.
AIMS: Non-laser-based methods are safe in lead extraction but in the past have been less effective than laser methods. In the past decade, new equipment has been introduced including the Evolution® Mechanical Dilator Sheath and the Evolution® RL. We sought to determine the impact of new equipment on outcome in mechanical lead extraction. METHODS AND RESULTS: We considered 288 consecutive patients (age 66 ± 18 years) who underwent transvenous lead extraction (TLE) of 522 leads in the decade to the end of 2014. Three groups were identified: Group 1 (pre-Evolution® period, 76 patients, 133 leads), Group 2 (original Evolution® period, 115 patients, 221 leads), and Group 3 (Evolution® RL period, 97 patients, 168 leads). The age of leads was significantly greater in Groups 2 and 3 (6.2 ± 4.4 and 6.1 ± 5.4 years vs.4.7 ± 4.5, P < 0.05) as was the proportion of implantable cardioverter defibrillator leads (27.2 and 28.9 vs. 14.3%, P < 0.05). The groups were similar in the number of leads extracted per patient. Despite the increasing complexity of the systems extracted, complete extraction was achieved in a progressively greater proportion of leads (88.0% in Group 1, 95.5% in Group 2, and 97.6% in Group 3, P < 0.05), and procedure duration was similar. The proportion of leads for which femoral access was required was greater in Group 3 (11%, 18/164) compared with Group 2 (3%, 7/211), P = 0.006. The only major complications were a post-procedure subacute tamponade in Group 1 and an oesophageal injury related to transoesophageal echocardiography in Group 3. CONCLUSION: With current equipment, mechanical extraction provides a good combination of efficacy and safety.
No takes yet. Share an insight, caveat, or question.
Domenichini et al. (2017) conducted a cohort in Transvenous lead extraction (n=288). Evolution and Evolution RL mechanical extraction sheaths vs. Pre-Evolution period equipment was evaluated on Complete lead extraction (p=< 0.05). Newer mechanical extraction equipment (Evolution and Evolution RL) progressively increased the rate of complete lead extraction from 88.0% to 97.6%.
Synapse has enriched 10 related papers. Consider them for comparative context: