Why the study?
Guidelines recommend hs-cTn methods for detecting myocardial injury in suspected AMI, prompting verification and validation of the new Mindray hs-cTnI assay on the CL2600i analyzer compared to the Abbott Alinity-i series instrument.
Does the Mindray hs-cTnI assay on the CL2600i analyzer demonstrate comparable analytical performance to the Abbott Alinity-i system?
Does the Mindray hs-cTnI assay on the CL2600i analyzer demonstrate comparable analytical performance to the Abbott Alinity-i system?
The new Mindray hs-cTnI assay demonstrates robust analytical performance comparable to established Abbott assays, supporting its clinical utility for diagnosing acute myocardial infarction.
New Mindray cTn assay meets analytical verification; leaves open clinical performance for AMI diagnosis in ED settings.
Background: International guidelines recommend the use of high-sensitivity cardiac troponin (hs-cTn) I and T methods for the detection of myocardial injury as a pre-requisite for the diagnosis of acute myocardial infarction (AMI) in patients admitted to the emergency department. Recently, Mindray (Mindray Bio-Medical Electronics Co., Ltd., Shenzhen, China) has introduced a new chemiluminescence immunoassay (CLIA) for the detection of the cTn complex. The present study aims to verify and validate the hs-cTnI Mindray assay on the new automated CL2600i analyzer compared to the routine Alinity-i series instrument by Abbott (Abbott, Chicago, IL, USA). Methods: This study evaluated linearity, precision through the 5 × 5 protocol, methodological comparison on plasma and serum matrices, hs-cTnI 99th percentile imprecision, and the hs-cTnI detection rate in a healthy population. Results: The results obtained proved that the performance of the Mindray hs-cTnI test on the CL2600i platform was closely comparable to the Abbott Alinity-i system (plasma R2: 0.974; serum R2: 0.995). The CVs were consistently low, and no significant differences were reported. Excellent analytical performance, with high sensitivity, was also observed in the healthy population (overall detection rate: 79%), as well as good linearity within the measuring range (R2: 0.994). Conclusions: The Mindray hs-cTnI test confirms its robustness and utility in routine practice as an advanced assay. The new technology, with more sensitive detection methods, may improve the accuracy and reliability of cardiac biomarker testing, ultimately leading to better outcomes in the management of patients with AMI and other cardiac conditions.
No takes yet. Share an insight, caveat, or question.
Tomassetti et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: