Key result
Multislice MRI tracks perfusion defects with ~85% agreement against scintigraphy and detects post-revascularization gains.
Observational (n=11)
Answers are grounded in this paper and cite their sources.
Demonstrates multislice MRI accurately tracks myocardial perfusion recovery after revascularization in coronary artery disease, highlighting its utility for evaluating therapeutic success.
BACKGROUND: Our purpose was to use multislice MRI for detection of reversible myocardial ischemia and assessment of the effect of revascularization on tissue perfusion in patients with coronary artery disease. METHODS AND RESULTS: Eleven patients with single-vessel proximal left anterior descending coronary artery disease were studied with MRI and thallium scintigraphy before and 3 months after revascularization. All patients had a reversible perfusion defect by scintigraphy before treatment. With a 1.5-T MR imager, IR-prepared turboflash images were acquired in three left ventricular short-axis planes during 0.05 mmol/kg Gd-DTPA bolus at rest and with dipyridamole-induced stress. Before treatment, stress increased enhancement slope in normal (6.4+/-4.4 to 7.4+/-5.0 s(-1), P<.04) and decreased it in underperfused (5.4+/-3.7 to 2.6+/-1.4 s(-1), P<.02) regions, resulting in a contrast-to-noise ratio of 6.87+/-3.09 in underperfused myocardium. Revascularization normalized enhancement patterns of the formerly underperfused myocardium and decreased defect size both in scintigraphy (66+/-53 degrees to 8+/-12 degrees, P<.001) and MRI sections (49+/-41 degrees to 9+/-8 degrees, P<.001). Agreement of 85% in detection and correlation of 0.86 (SEE, 21 degrees, P<.001) in sizing perfusion defects was found between MRI and scintigraphy. CONCLUSIONS: Multislice contrast-enhanced MRI can be used to detect myocardial perfusion defects in patients with coronary artery disease and in assessment of the effect of treatment on myocardial perfusion.
No takes yet. Share an insight, caveat, or question.
Lauerma et al. (1997) conducted an observational in Single-vessel proximal left anterior descending coronary artery disease (n=11). Multislice contrast-enhanced MRI (0.05 mmol/kg Gd-DTPA bolus at rest and dipyridamole stress) before and after revascularization vs. Thallium scintigraphy / pre-revascularization baseline was evaluated on Myocardial perfusion defect size and agreement with thallium scintigraphy (r = 0.86 (85% detection agreement), p=<0.001). Multislice MRI showed 85% agreement and r=0.86 with thallium scintigraphy in sizing perfusion defects, which decreased from 49° to 9° after revascularization.
Synapse has enriched 10 related papers. Consider them for comparative context: