Why the study?
It was unknown whether Platelet Function Analyzer PFA-100 and Plateletworks point-of-care devices provide comparable and useful platelet function information to manage anti-aggregated patients before surgery.
Does Plateletworks® correlate with PFA-100® for measuring platelet function before urgent surgery in patients on chronic antiplatelet therapy?
Does Plateletworks® correlate with PFA-100® for measuring platelet function before urgent surgery in patients on chronic antiplatelet therapy?
Answers are grounded in this paper and cite their sources.
There is a weak concordance between Plateletworks and PFA-100 for measuring platelet function in elderly patients on chronic antiplatelet therapy before urgent surgery.
Hemostasis is crucial for reducing bleeding during surgical procedures. The points-of-care based on the platelet function test could be useful to minimize the complications related to chronic antiplatelet therapy during surgery. The present study is aimed at comparing two point-of-care platelet function devices—Platelet Function Analyzer PFA-100® (Siemens Canada, Mississauga, ON, Canada) and Plateletworks®(Helena Laboratories, Beaumont, TX, USA). Our objective is to evaluate if they provide comparable and useful information to manage anti-aggregate patients before surgery. We included patients with a femoral fracture receiving chronic antiplatelet therapy and a median age of 89 years (range from 70 to 98). A platelet function evaluation was performed on all patients before surgery using both devices—Plateletworks® and PFA-100®. The correlation between Plateletworks® and PFA-100® was performed using Cohen’s Kappa coefficient. Twenty consecutive patients participated in the trial; 16 patients were under treatment with 75 mg/day of clopidogrel, three with >300 mg/day of acetylsalicylic acid (ASA), and only one was in treatment with both antiplatelet agents. Cohen’s Kappa coefficient was 0.327 comparing PFA-100®-ADP (adenosine diphosphate) and Plateletworks® and, 0.200 comparing PFA-100®-EPI (epinephrine) and Plateletworks®. In conclusion, we found a weak concordance comparing PFA-100® and Plateletworks®. This could partially be due to the advanced age of the included patients. However, given the limited sample size, more studies are necessary to confirm these results.
No takes yet. Share an insight, caveat, or question.
Anaya et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: