Answers are grounded in this paper and cite their sources.
A new detection system for measurement of accurate saturation transfer electron paramagnetic resonance (ST-EPR) spectra has been developed. The principle of this system is that the EPR absorbance is recorded digitally over a wide frequency range without filtering, then processed numerically by a Fourier transformation technique without any phase sensitive detector which is used in conventional ST-EPR spectroscopy. It is possible to acquire the first and second harmonic signals at in-phase and 90° out-of-phase simultaneously. To obtain very high accuracy of phase determination, the modulation signal is also recorded and processed digitally simultaneously. To improve the signal/noise ratio, an accumulation technique is adopted. Consequently, in-phase and 90° out-of-phase for 100 kHz modulation can be easily corrected within the ±0.5° accuracy, and the phase can be changed quantitatively. Experimentally it is shown that 90° out-of-phase has to be set within 90 ± 0.5° to suppress the out-of-phase signal under 1% of in-phase signal in the absence of saturation. The errors in the ST-EPR parameters and τ-values are also estimated in the case of improper phase angles.
No takes yet. Share an insight, caveat, or question.
Watanabe et al. (1980) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: