Why the study?
Does the RBF neural network algorithm improve precision and process time compared to DD-HMM for recognizing ventricular tachycardia and fibrillation?
Does the RBF neural network algorithm improve precision and process time compared to DD-HMM for recognizing ventricular tachycardia and fibrillation?
Answers are grounded in this paper and cite their sources.
RBF neural networks provide higher precision but slower processing times compared to DD-HMM for automated recognition of shockable ventricular arrhythmias.
This paper deals with automatic recognition of cardiac arrhythmias that require immediate electrical defibrillation therapy (ventricular fibrillation and ventricular tachycardia), through ECG (electrocardiogram) samples. The DD-HMM (discrete density hidden Markov model) and RBF (radial basis function) neural network algorithms were compared in the following aspects: precision, defined as correct recognition percentage and process time, defined as the delay since the ECG input until the result, indicating shock or non-shock events. The results show that RBF is more precise than DD-HMM but not so fast to evaluate. PhysioNet database files were used to train and to validate the algorithms.
No takes yet. Share an insight, caveat, or question.
Scolari et al. (2007) studied this question.