Why the study?
Does the oscillometric method accurately measure blood pressure compared to the auscultatory method in hemodialysed patients?
Does the oscillometric method accurately measure blood pressure compared to the auscultatory method in hemodialysed patients?
The oscillometric method underestimates diastolic blood pressure compared to the auscultatory method in hemodialysis patients, with individual patient features rather than the hemodialysis process itself determining measurement accuracy.
Clinicians should interpret oscillometric DBP cautiously in hemodialysis patients; this Level 4 finding leaves open outcome impact and needs prospective validation.
BACKGROUND: Despite more frequent use of the oscillometric method (OSC) for arterial blood pressure (BP) measurement, little is known about OSC's accuracy when used with hemodialysed patients. This study was undertaken to determine if hemodialysis (HD) and individual features in examined patients can affect the accuracy of OSC for BP measurement. MATERIAL/METHODS: In 54 hemodialysed patients (57 ± 15 years), during 2 sessions (before and after HD), 3 pairs of BP measurements each were performed on arms, alternately employing OSC and auscultatory method, with mercury manometers by 2 observers (REF). RESULTS: No difference was found in systolic BP measured before and after HD (SBP before HD: REF -147.3 ± 27.3, OSC -147.5 ± 25.0 mmHg, p>0.05, SBP after HD: REF-141.1 ± 33.9, OSC-141.2 ± 31.5 mmHg, p>0.05, respectively), but diastolic BP (DBP) was significantly higher both before and after HD during REF measurement in comparison with OSC (DBP before HD: REF -79.0 ± 17.0, OSC -76.7 ± 15.0 mmHg, DBP after HD: REF -78.6 ± 18.8, OSC -76.7 ± 16.7 mmHg, p<0.001, respectively). No significant correlation between loss of body weight caused by HD and differences in BP measured by REF and OSC after HD was indicated (Pearson's correlation coefficients: for SBP -0.041, for DBP 0.030). However, a significant correlation between differences in BP measured by REF and OSC before HD and differences in BP measured by REF and OSC after HD was observed (Spearman's rank correlation coefficients: for SBP 0.502 and for DBP 0.557, p<0.000001). CONCLUSIONS: Our study found that individual features, not HD, determine the accuracy of OSC for BP measurement in hemodialysed patients.
No takes yet. Share an insight, caveat, or question.
Czarkowski et al. (2011) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: