Key result
The HD-BPM automated oscillometric sphygmomanometer yielded mean differences of 0.5 +/- 9.0 mmHg (systolic) and 1.5 +/- 6.9 mmHg (diastolic) compared to auscultatory measurements during dialysis.
Why the study?
Does the HD-BPM automated oscillometric sphygmomanometer provide accurate blood pressure measurements compared to standard methods in dialysis patients?
Observational (n=53)
Yes
Does the HD-BPM automated oscillometric sphygmomanometer provide accurate blood pressure measurements compared to standard methods in dialysis patients?
The HD-BPM automated oscillometric sphygmomanometer provides acceptable blood pressure measurement accuracy during haemodialysis, highlighting the importance of evaluating device accuracy in specific clinical settings.
Answers are grounded in this paper and cite their sources.
May support HD-BPM deployment in dialysis units; extends validation data to intradialytic settings but leaves prospective outcome studies open.
BACKGROUND: Nowadays, an increasing number of dialysis machines lodge a blood pressure (BP) measuring device, whose accuracy has a clear implication for the patients' clinical management. METHODS: An automated oscillometric sphygmomanometer (HD-BPM by Gambro Dasco) used during haemodialysis was submitted to both in vitro and in vivo tests, in order to evaluate some modifications aimed at improving measurement accuracy and consistency. The results were compared with those obtained by another oscillometric monitor (BX-100 by Colin). Three steps of evaluation were followed. First, the maintenance of the overall accuracy requirements prescribed by ANSI/AAMI SP-10 standard was verified. Then, an in vitro validation was carried out by using a test simulator. Finally, during a multi-centre field trial, 392 BP measurement sessions on 53 dialysis patients were collected. Every session consisted of two consecutive intra-dialysis measurements by the oscillometric monitors, each one performed simultaneously to an auscultatory measurement. A comparison with the intra-arterial method was performed as well. RESULTS: When compared with an in vivo data set previously collected, the HD-BPM accuracy complied with required limits. Second, the internal repeatability with respect to the simulator was satisfactory (SD of the differences between device and simulator readings: HD-BPM: systolic = 5.7, diastolic = 4.2; BX-100: systolic = 4.2, diastolic = 5.5 mmHg). Moreover, the comparison between oscillometric and auscultatory methods during in vivo trial gave similar results for the two monitors, even if systolic pressure SD exceeds the limit recommended by ANSI/AAMI SP-10 (mean value of the differences +/- SD: HD-BPM: systolic = 0.5 +/- 9.0, diastolic = 1.5 +/- 6.9; BX-100: systolic = 3.1 +/- 8.2, diastolic = -2.0 +/- 7.6 mmHg). CONCLUSIONS: These data underline the importance of performing accuracy evaluations for BP monitors in the conditions where they normally work, by using well-accepted protocols.
No takes yet. Share an insight, caveat, or question.
Lodi et al. (2007) conducted an observational in Dialysis (n=53). Automated oscillometric sphygmomanometer (HD-BPM) vs. Auscultatory measurement and BX-100 monitor was evaluated on Accuracy of blood pressure measurement compared to auscultatory method. The HD-BPM automated oscillometric sphygmomanometer yielded mean differences of 0.5 +/- 9.0 mmHg (systolic) and 1.5 +/- 6.9 mmHg (diastolic) compared to auscultatory measurements during dialysis.