Why the study?
Does the Mobil-O-Graph oscillometric device provide comparable estimates of central systolic pressure and arterial stiffness compared to the Sphygmocor tonometric device in hemodialysis patients?
Does the Mobil-O-Graph oscillometric device provide comparable estimates of central systolic pressure and arterial stiffness compared to the Sphygmocor tonometric device in hemodialysis patients?
The Mobil-O-Graph oscillometric device provides acceptable estimates of central systolic blood pressure and augmentation index, but slightly underestimates pulse wave velocity compared to tonometry in hemodialysis patients.
Answers are grounded in this paper and cite their sources.
May support Mobil-O-Graph for aSBP and AIx in hemodialysis; leaves open PWV accuracy versus tonometry.
BACKGROUND/AIMS: Elevated wave reflections and arterial stiffness, as well as ambulatory blood pressure (BP) are independent predictors of cardiovascular risk in end-stage-renal-disease. This study is the first to evaluate in hemodialysis patients the validity of a new ambulatory oscillometric device (Mobil-O-Graph, IEM, Germany), which estimates aortic BP, augmentation index (AIx) and pulse wave velocity (PWV). METHODS: Aortic SBP (aSBP), heart rate-adjusted AIx (AIx(75)) and PWV measured with Mobil-O-Graph were compared with the values from the most widely used tonometric device (Sphygmocor, ArtCor, Australia) in 73 hemodialysis patients. Measurements were made in a randomized order after 10 min of rest in the supine position at least 30 min before a dialysis session. Brachial BP (mercury sphygmomanometer) was used for the calibration of Sphygmocor's waveform. RESULTS: Sphygmocor-derived aSBP and AIx(75) did not differ from the relevant Mobil-O-Graph measurements (aSBP: 136.3 ± 19.6 vs. 133.5 ± 19.3 mm Hg, p = 0.068; AIx(75): 28.4 ± 9.3 vs. 30.0 ± 11.8%, p = 0.229). The small difference in aSBP is perhaps explained by a relevant difference in brachial SBP used for calibration (146.9 ± 20.4 vs. 145.2 ± 19.9 mm Hg, p = 0.341). Sphygmocor PWV was higher than Mobil-O-Graph PWV (10.3 ± 3.4 vs. 9.5 ± 2.1 m/s, p < 0.01). All 3 parameters estimated by Mobil-O-Graph showed highly significant (p < 0.001) correlations with the relevant measurements of Sphygmocor (aSBP, r = 0.770; AIx(75), r = 0.400; PWV, r = 0.739). The Bland-Altman Plots for aSBP and AIx(75) showed acceptable agreement between the two devices and no evidence of systemic bias for PWV. CONCLUSION: As in other populations, acceptable agreement between Mobil-O-Graph and Sphygmocor was evident for aSBP and AIx(75) in hemodialysis patients; PWV was slightly underestimated by Mobil-O-Graph.
No takes yet. Share an insight, caveat, or question.
Sarafidis et al. (2014) studied this question.