Key result
The Arteriograph device showed no significant correlation with the validated PulsePen tonometer for measuring pulse wave velocity and lacked prognostic value for CV mortality.
Why the study?
Does the Arteriograph device provide valid measurements of arterial stiffness compared to the PulsePen tonometer in patients on maintenance hemodialysis?
Observational (n=98)
Does the Arteriograph device provide valid measurements of arterial stiffness compared to the PulsePen tonometer in patients on maintenance hemodialysis?
The Arteriograph device has limited validity for determining pulse wave velocity and predicting cardiovascular mortality in hemodialysis patients compared to the validated PulsePen tonometer.
Answers are grounded in this paper and cite their sources.
Arteriograph lacks reliability for PWV assessment in hemodialysis; leaves open need for device validation before any clinical use.
Background: Measuring arterial stiffness (augmentation index (AI), aortic pulse wave velocity (PWV)) in hemodialysis (HD) patients has prognostic significance. To assess its validity, the new oscillometric Arteriograph device (AIA, PWVA) was compared to the validated PulsePen tonometer (AIP, PWVP). Methods: AI and PWV were measured in 98 patients with both devices before HD. Validity was evaluated by Pearson’s correlation, Bland-Altman analysis, and by assessing the prognostic value of AI and PWV to predict cardiovascular (CV) mortality over 29 months. Results: Correlation between AIP and AIA was significant (R = 0.527, p < 0.001). The mean difference of AI values obtained by the two devices was –20.6%, and 30% of the paired AI differences fall outside the ±1 SD boundary of the mean between-device difference. There was no significant correlation between the PWVP and PWVA readings (R = 0.173, p = 0.097). The average difference of PWV values by the two devices was –1.2 m/s, and 20.6% of the paired PWV differences fall outside the ±1 SD boundary. In survival analyses, only PWVP but not PWVA was significantly related to CV mortality. Conclusion: Lack of correlation between PWVP and PWVA and lack of prognostic significance of PWVA suggest limited validity of Arteriograph to determine PWV in patients on HD.
No takes yet. Share an insight, caveat, or question.
Nemcsik et al. (2009) conducted an observational in Maintenance hemodialysis (n=98). Arteriograph device vs. PulsePen tonometer was evaluated on Correlation of pulse wave velocity (PWV) measurements between devices (R 0.173, p=0.097). The Arteriograph device showed no significant correlation with the validated PulsePen tonometer for measuring pulse wave velocity and lacked prognostic value for CV mortality.