Answers are grounded in this paper and cite their sources.
Purpose: We have developed an infrared earphone-type thermometer (“CE-thermo®”) able to continuously monitor body temperature in pediatric patients undergoing general anesthesia. In this study, we evaluated the usefulness of this device in pediatric patients receiving general anesthesia for non-cardiac surgery.Methods: After approval from Sapporo Medical University Research Ethics Board, twenty pediatric patients scheduled for non-cardiac surgery under general anaesthesia were enrolled in this study. During general anesthesia, both esophageal temperature (Teso) and tympanic temperature (Ttym, measured by “CE-thermo®”) were monitored and continuously recorded for subsequent analysis. This is a pilot study that enrolled twenty patients scheduled for non-cardiac surgery under general anaesthesia. All patients were ASA physical status 1 or 2 and aged between 4 and 12 years.Results: The Pearson’s correlation coefficient was 0.84, indicating a strong correlation between oesophageal temperature and tympanic temperature. A Bland-Altman plot showed that the mean bias of the “CE-thermo®” was -0.19, below that of Teso (2 SD ± 0.60°C).Conclusion: The new earphone-type thermometer “CE-thermo®” is highly reliable for core body temperature monitoring in pediatric patients during general anesthesia.
No takes yet. Share an insight, caveat, or question.
Tachibana et al. (2017) studied this question.