Does a quadrature surface coil improve signal-to-noise ratio compared to a standard phased-array coil in carotid artery imaging?
Does a quadrature surface coil improve signal-to-noise ratio compared to a standard phased-array coil in carotid artery imaging?
A custom-built quadrature surface coil provides better signal-to-noise ratio for carotid artery imaging compared to standard phased-array coils.
Answers are grounded in this paper and cite their sources.
Should not yet alter coil selection for carotid MRI; leaves open in vivo validation of SNR gains.
PURPOSE: To evaluate the ability of a custom-built coil design to provide improved signal-to-noise ratio (SNR) and less signal drop with increasing depth at the carotid artery. MATERIALS AND METHODS: Phased-array surface coils can provide a high SNR to image the carotid vessel wall. However, given the required field-of-view (FOV) and penetration depth, these coils show either a fast signal drop with increasing depth or a moderate SNR at increased coil size. A quadrature surface coil (a butterfly coil in conjunction with a linear single-loop coil) was compared with a phased-array coil in phantom and human studies. RESULTS: The phantom studies showed that the quadrature coil has better SNR over the required FOV than a standard phased-array coil (26% at 3 cm depth). CONCLUSION: The quadrature coil enables better image quality to be achieved.