Why the study?
Does an expandable intravenous RF coil improve signal-to-noise ratio for arterial wall imaging compared to external coils in a phantom and porcine model?
Does an expandable intravenous RF coil improve signal-to-noise ratio for arterial wall imaging compared to external coils in a phantom and porcine model?
Answers are grounded in this paper and cite their sources.
An expandable intravenous RF coil provides a 15- to 20-fold increase in SNR compared to external coils, offering a promising alternative to intraarterial coils for high-resolution arterial wall imaging.
An intravenous (iv) radiofrequency (RF) coil is proposed as a means of obtaining high resolution images of artery wall. The anatomic positioning of peripheral artery/vein pairs was investigated and a phantom mimicking the iliac artery/vein pair was constructed. Imaging results, comparing iv coils with external coils, demonstrated a potential 15- to 20-fold increase in signal-to-noise ratio (SNR) with iv coils. The SNR benefit was measured over a cylindrical volume, adjacent to the coil, and typical of artery position. Prototype expandable iv coils were constructed of Cu-Be loops and introduced via an 8-Fr catheter. The effects of local and remote iv coil tuning were investigated and local tuning was found to provide significant SNR benefits. The in vivo performance of iv RF coils was demonstrated in a porcine animal model. The iv coils were found to be an excellent alternative to intraarterial coils.
No takes yet. Share an insight, caveat, or question.
Martin et al. (1998) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: