High on-treatment platelet reactivity is linked to stroke recurrence, making it important to find a reliable assessment method.
What is the correlation between VerifyNow, TEG, and PL-12 in assessing high on-treatment platelet reactivity in patients with minor ischemic stroke or TIA?
What is the correlation between VerifyNow, TEG, and PL-12 in assessing high on-treatment platelet reactivity in patients with minor ischemic stroke or TIA?
Answers are grounded in this paper and cite their sources.
VerifyNow and TEG are more correlated than PL-12 for assessing ADP-induced high on-treatment platelet reactivity, though PL-12 identified patients with higher rates of stroke recurrence.
High on-treatment platelet reactivity (HOPR) is associated with stroke recurrence. It is important to find a reliable method to assess HOPR. We aimed to compare the correlations between VerifyNow™ system, thromboelastography (TEG), and Aggrestar platelet function analyzer (PL-12) on platelet activity in patients with minor ischemic stroke or transient ischemic attack (TIA) after dual antiplatelet therapy for 7 days. About 276 patients were included. Spearman's correlation coefficient and the kappa coefficient were adopted to evaluate associations among the three test methods. An obvious correlation between VerifyNow and TEG on HOPR-ADP (r=0.64, p<0.001) was found. The correlations of HOPR-ADP between PL-12 and the other two platelet function analyzers were moderate (PL-12 versus VerifyNow, r=0.47, p<0.001; PL-12 versus TEG, r=0.25, p<0.001). The correlations of HOPR-AA were limited among these three platelet function analyzers (VerifyNow versus TEG, r=0.09, p=0.14; VerifyNow versus PL-12, r=0.15, p=0.01; PL-12 versus TEG, r=0.10, p=0.09). Correlations among different platelet function analyzers were varied. VerifyNow and TEG were more correlative than PL-12 on HOPR-ADP. The consistence of HOPR-AA was limited among VerifyNow, TEG and PL-12. The proportion of stroke recurrence and composite events in patients with HOPR-ADP assessing by PL-12 was higher than VerifyNow and TEG.
No takes yet. Share an insight, caveat, or question.
Ma et al. (2021) studied this question.