Answers are grounded in this paper and cite their sources.
A high-performance liquid chromatographic (HPLC) method for determining verapamil and norverapamil in plasma is presented and compared with gas chromatography/mass spectrometry (GC-MS). The plasma samples were extracted at alkaline pH with hexane containing 2-butanol (20 mL/L) and then back-extracted into phosphate buffer (0.1 mol/L, pH 3.0). For chromatography we used a reversed-phase column (Supelcosil LC-18 DB) with a mobile phase of the phosphate buffer and acetonitrile (70/30 by vol). Fluorescence detection was used (excitation at 203 nm, emission at 320 nm). Overall analytical recovery was 85%. Standard curves were linear from 1 to 1000 micrograms/L. The detection limit was 1 microgram/L. The assays are accurate and precise. We found no interferences by those substances tested. Results by HPLC and GC-MS agreed well (r = 0.99) for both verapamil and norverapamil determinations.
No takes yet. Share an insight, caveat, or question.
Hynning et al. (1988) studied this question.