Cryo-electron tomography (CET) is a well-established technique for imaging cellular and molecular structures at sub-nanometer resolution. As the method is limited to samples that are thinner than 500 nm, suitable sample preparation is required to attain CET data from larger cell volumes. Recently, cryo-focused ion beam (cryo-FIB) milling of plunge-frozen biological material has been shown to reproducibly yield large, homogeneously thin, distortion-free vitreous cross-sections for state-of-the-art CET. All eukaryotic and prokaryotic cells that can be plunge-frozen can be thinned with the cryo-FIB technique. Together with advances in low-dose microscopy, this has shifted the frontiers of in situ structural biology. In this protocol we describe the typical steps of the cryo-FIB technique, starting with fully grown cell cultures. Three recently investigated biological samples are given as examples.
No takes yet. Share an insight, caveat, or question.
Schaffer et al. (2015) studied this question.
Synapse has enriched 3 closely related papers on similar clinical questions. Consider them for comparative context: