Key result
Contact force-sensing catheters show no long-term success benefit over standard catheters for outflow tract VA ablation.
Why the study?
Adequate catheter-tissue contact is important for optimal lesion formation, but the potential benefits of contact force-sensing catheters during non-fluoroscopic RFCA of outflow tract ventricular arrhythmias needed evaluation.
Does the use of contact force-sensing catheters improve acute and long-term procedural success or safety compared to standard irrigated catheters in patients undergoing non-fluoroscopic radiofrequency catheter ablation for idiopathic outflow tract ventricular arrhythmias?
Cohort (n=102)
Does the use of contact force-sensing catheters improve acute and long-term procedural success or safety compared to standard irrigated catheters in patients undergoing non-fluoroscopic radiofrequency catheter ablation for idiopathic outflow tract ventricular arrhythmias?
Answers are grounded in this paper and cite their sources.
No added benefit from contact force-sensing catheters; leaves open their role in non-fluoroscopic outflow tract VA ablation.
Background: Adequate contact between the catheter tip and tissue is important for optimal lesion formation and, in some procedures, it has been associated with improved effectiveness and safety. We evaluated the potential benefits of contact force-sensing (CFS) catheters during non-fluoroscopic radiofrequency catheter ablation (NF-RFCA) of idiopathic ventricular arrhythmias (VAs) originating from outflow tracts (OTs). Methods: A group of 102 patients who underwent NF-RFCA (CARTO, Biosense Webster Inc., Irvine, CA, USA) of VAs from OTs between 2014 to 2018 was retrospectively analyzed. Results: We included 52 (50.9%) patients in whom NF-RFCA was performed using CFS catheters and 50 (49.1%) who were ablated using standard catheters. Arrhythmias were localized in the right and left OT in 70 (68.6%) and 32 (31.4%) patients, respectively. The RFCA acute success rate was 96.1% (n = 98) and long-term success during a minimum 12-month follow-up (mean 51.3 ± 21.6 months) was 85.3% (n = 87), with no difference between CFS and standard catheters. There was no difference in complications rate between CFS (n = 1) and standard catheter (n = 2) ablations. Conclusions: There is no additional advantage of CFS catheters use over standard catheters during NF-RFCA of OT-VAs in terms of procedural effectiveness and safety.
No takes yet. Share an insight, caveat, or question.
Karkowski et al. (2022) conducted a cohort in Idiopathic outflow tract ventricular arrhythmias (OT-VAs) (n=102). Contact force-sensing (CFS) open-irrigated catheter ablation vs. Standard open-irrigated catheter ablation without contact force-sensing was evaluated on Long-term ablation success (over 80% reduction of arrhythmia burden after at least 3 months in repeated 24-h ECG monitoring during a minimum 12-month follow-up) (p=0.357). Contact force-sensing catheters showed no significant difference in long-term success compared to standard catheters (88.5% vs. 82.0%, p=0.357) during non-fluoroscopic ablation of outflow tract VAs.