Key result
A combination of three measurements with the MyDiagnostick device demonstrated a sensitivity of 94% and a specificity of 93% for detecting atrial fibrillation in primary care.
Why the study?
Does the MyDiagnostick device accurately detect atrial fibrillation compared to a 12-lead ECG in primary care patients?
Cross-Sectional (n=191)
Single-blind
Yes
Does the MyDiagnostick device accurately detect atrial fibrillation compared to a 12-lead ECG in primary care patients?
The MyDiagnostick device demonstrates high sensitivity and specificity for detecting atrial fibrillation in primary care, offering a potential alternative to pulse palpation for screening.
Answers are grounded in this paper and cite their sources.
May aid primary care AF screening; extends device accuracy data but leaves open outcome validation.
BACKGROUND: Atrial fibrillation is very common in people aged 65 or older. This condition increases the risk of death, congestive heart failure and thromboembolic conditions. Many patients with atrial fibrillation are asymptomatic and a cerebrovascular accident (CVA) is often the first clinical presentation. Guidelines concerning the prevention of CVA recommend monitoring the heart rate in patients aged 65 or older. Recently, the MyDiagnostick (Applied Biomedical Systems BV, Maastricht, The Netherlands) was introduced as a new screening tool which might serve as an alternative for the less accurate pulse palpation. This study was designed to explore the diagnostic accuracy of the MyDiagnostick for the detection of atrial fibrillation. METHODS: A phase II diagnostic accuracy study in a convenience sample of 191 subjects recruited in primary care. The majority of participants were patients with a known history of atrial fibrillation (n=161). Readings of the MyDiagnostick were compared with electrocardiographic recordings. Sensitivity and specificity and their 95% confidence interval were calculated using 2x2 tables. RESULTS: A prevalence of 54% for an atrial fibrillation rhythm was found in the study population at the moment of the study. A combination of three measurements with the MyDiagnostick for each patient showed a sensitivity of 94% (95% CI 87 - 98) and a specificity of 93% (95% CI 85 - 97). CONCLUSION: The MyDiagnostick is an easy-to-use device that showed a good diagnostic accuracy with a high sensitivity and specificity for atrial fibrillation in a convenience sample in primary care. Future research is needed to determine the place of the MyDiagnostick in possible screening or case-finding strategies for atrial fibrillation.
No takes yet. Share an insight, caveat, or question.
Vaes et al. (2014) conducted a cross-sectional in Atrial fibrillation (n=191). MyDiagnostick vs. 12-lead ECG was evaluated on Diagnostic accuracy (sensitivity and specificity) for detecting atrial fibrillation (Sensitivity 94%, Specificity 93%, 95% CI 87-98). A combination of three measurements with the MyDiagnostick device demonstrated a sensitivity of 94% and a specificity of 93% for detecting atrial fibrillation in primary care.