Answers are grounded in this paper and cite their sources.
Factory programmed pacemaker features like + PVARP on PVC can cause cycle length variability and trigger recurrent polymorphic ventricular tachycardia in susceptible patients.
A patient with refractory bradycardia dependent polymorphic ventricular tachycardia (PMVT) underwent permanent pacemaker placement for suppression of her arrhythmia. Following implant, the patient had recurrent PMVT. Interrogation of the device revealed a factory programmed feature (+ PVARP on PVC) designed to prevent pacemaker-mediated tachycardia that led to cycle length variability and subsequent arrhythmia recurrence. This feature was deactivated and the patients' arrhythmia did not recur.
No takes yet. Share an insight, caveat, or question.
Goldman et al. (1998) studied this question.