Why the study?
Does the non-invasive cardiac output system (NICO) provide reliable cardiac output measurements compared to a pulmonary artery catheter in patients undergoing off-pump coronary surgery?
Does the non-invasive cardiac output system (NICO) provide reliable cardiac output measurements compared to a pulmonary artery catheter in patients undergoing off-pump coronary surgery?
Answers are grounded in this paper and cite their sources.
The non-invasive cardiac output system provides reliable and rapid cardiac output measurements compared to a pulmonary artery catheter during off-pump coronary surgery.
BACKGROUND: The aim of this prospective study was to compare continuous cardiac output measurements of the non-invasive cardiac output system (NICO) with the pulmonary artery catheter during off-pump coronary bypass surgery. METHODS: Twenty-two patients enrolled for off-pump coronary surgery received both a pulmonary artery catheter and a non-invasive cardiac output system for measurement of cardiac output. Data were compared by the Bland-Altman method to calculate the degree of agreement and to analyse if a significant difference existed between the two methods of cardiac output measurements. RESULTS: Perioperatively, the non-invasive cardiac output underestimated cardiac output, but postoperatively overestimated it. The limits of agreement were larger during surgery compared to the postoperative period (-3.1; +2.5 vs. -1.4; +2.2 L min(-1)). Perioperatively, cardiac output measured with the pulmonary artery catheter varied from 0.5 to 7.5 L min(-1) (mean 3.6 L min(-1)) and with the non-invasive cardiac output from 0.5 to 8.4 L min(-1) (mean 3.9 L min(-1)). Postoperatively, these were 2.5-7.7 L min(-1) (mean 4.5 L min(-1)) and 2.3-8.4 L min(-1) (mean 4.9 L min(-1)), respectively. CONCLUSION: During off-pump cardiac surgery, the non-invasive cardiac output reliably measures cardiac output and does it more rapidly than a pulmonary artery catheter and may be more useful in order to detect rapid haemodynamic changes.
No takes yet. Share an insight, caveat, or question.
Guéret et al. (2006) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: