Why the study?
Does the FloTrac/Vigileo system accurately measure cardiac output compared to a pulmonary artery catheter in patients undergoing off-pump coronary bypass surgery with reduced LVEF?
Does the FloTrac/Vigileo system accurately measure cardiac output compared to a pulmonary artery catheter in patients undergoing off-pump coronary bypass surgery with reduced LVEF?
The FloTrac/Vigileo system provides reliable cardiac output measurements during off-pump coronary bypass surgery, even in patients with reduced LVEF, provided systemic vascular resistance is within normal limits.
Answers are grounded in this paper and cite their sources.
Supports FloTrac/Vigileo CO monitoring in reduced-LVEF off-pump CABG when SVR normal; extends observational validation but leaves open need for RCTs.
BACKGROUND: We compared the continuous cardiac index measured by the FloTrac/Vigileo™ system (FCI) to that measured by a pulmonary artery catheter (CCI) with emphasis on the accuracy of the FCI in patients with a decreased left ventricular ejection fraction (LVEF) and a low cardiac output status during off-pump coronary bypass surgery (OPCAB). We also assessed the influence of several factors affecting the pulse contour, such as the mean arterial pressure (MAP), the systemic vascular resistance index (SVRI) and the use of norepinephrine. METHODS: Fifty patients who were undergoing OPCAB (30 patients with a LVEF ≥ 40%, 20 patients with a LVEF < 40%) were enrolled. The FCI and CCI were measured and we performed a Bland-Altman analysis. Subgroup analyses were done according to the LVEF (< 40%), the CCI (≤ 2.4 L/min/m), the MAP (60-80 mmHg), the SVRI (1,600-2,600 dyne/s/cm(5)/m(2)) and the use of norepinephrine. RESULTS: The FCI was reliable at all the time points of measurement with an overall bias and limit of agreement of -0.07 and 0.67 L/min/m(2), respectively, resulting in a percentage error of 26.9%. The percentage errors in the patients with a decreased LVEF and in a low cardiac output status were 28.2% and 22.3%, respectively. However, the percentage error in the 91 data pairs outside the normal range of the SVRI was 40.2%. CONCLUSIONS: The cardiac output measured by the FloTrac/Vigileo™ system was reliable even in patients with a decreased LVEF and in a low cardiac output status during OPCAB. Acceptable agreement was also noted during the period of heart displacement and grafting of the obtuse marginalis branch.
No takes yet. Share an insight, caveat, or question.
Jo et al. (2011) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: