Why the study?
The reproducibility of longitudinal strain measured by 2D speckle tracking echocardiography may be affected by ultrasound settings, prompting evaluation of the effect of transmit ultrasound frequency on global longitudinal strain.
Does ultrasound transmit frequency affect the measurement of global longitudinal strain on 2D speckle tracking echocardiography?
Does ultrasound transmit frequency affect the measurement of global longitudinal strain on 2D speckle tracking echocardiography?
Ultrasound transmit frequency has a modest but statistically significant effect on global longitudinal strain measurements, though the absolute difference of ~1% is unlikely to be clinically significant in most routine cases.
Answers are grounded in this paper and cite their sources.
Supports routine GLS use regardless of frequency preset; leaves open multi-vendor confirmation and serial-scan effects.
The reproducibility of longitudinal strain measured by 2D speckle tracking echocardiography (2DSTE) may be affected by ultrasound settings. This study investigated the effect of transmit ultrasound frequency on global longitudinal strain (GLS) by 2DSTE. Apical, 2- and 4-chamber, and long-axis views were obtained in a consecutive group of 162 patients using Philips ultrasound devices. Three different frequency presets were used sequentially: high resolution (HRES, 1.9 to 2.1 MHz), general (HGEN, 1.6 to 1.8 MHz), and penetration mode (HPEN, 1.3 to 1.6 MHz). GLS values were determined for each preset using the Philips Q-station software, resulting in GLS-HRES, GLS-HGEN, and GLS-HPEN. Among the 151 patients with successfully measured GLS, a significant difference in GLS was observed among the three presets (p < 0.0001). GLS-HRES (-17.9 ± 4.4%) showed a slightly smaller magnitude compared to GLS-HGEN (-18.8 ± 4.5%, p < 0.0001) and GLS-HPEN (-18.8 ± 4.5%, p < 0.0001), with absolute differences of 1.1 ± 1.0% and 1.1 ± 1.2%, respectively. This variation in GLS with frequency was evident in patients with both optimal (n = 104) and suboptimal (n = 47) image quality and remained consistent regardless of ischemic etiology or ejection fraction. In conclusion, ultrasound frequency had only a modest effect on GLS measurements. GLS may be reliably assessed in most cases regardless of the ultrasound frequency used.