Answers are grounded in this paper and cite their sources.
Multi-turn split-conductor transmission-line resonators (MSTR) can be designed to operate over a wide range of NMR operating frequencies without lumped tuning capacitors. This architecture constitutes an alternative solution to recent designs proposed for high-Q, thin-film, high-temperature superconducting NMR probes. An advantage is that the resonant frequency can be calculated in a relatively simple manner in terms of coil turns or total coil length, coil width, substrate thickness, and dielectric constant. Analytical calculation of the resonant frequency is provided. Also, a series of MSTRs was constructed on a double copper-clad substrate, and their resonant frequencies are noted. The results obtained were in good agreement with the predicted values.
No takes yet. Share an insight, caveat, or question.
Serfaty et al. (1997) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: