Answers are grounded in this paper and cite their sources.
A simple procedure for the development of a range of polymeric ion-selective electrodes (ISEs) with low detection limits is presented. The electrodes were prepared by using a plasticizer-free methyl methacrylate-decyl methacrylate copolymer as membrane matrix and poly(3-octylthiophene) as intermediate layer deposited by solvent casting on gold sputtered copper electrodes as a solid inner contact. Five different electrodes were developed for Ag+, Pb2+, Ca2+, K+, and I-, with detection limits mostly in the nanomolar range. In this work, the lowest detection limits reported thus far with solid contact ISEs for the detection of silver (2.0 x 10(-9) M), potassium (10(-7) M), and iodide (10(-8) M) are presented. The developed electrodes exhibited a good response time and excellent reproducibility.
No takes yet. Share an insight, caveat, or question.
Chumbimuni‐Torres et al. (2006) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: