Online quantification of a single gas species over an extremely broad concentration range is important for complex industrial applications requiring both trace-level detection and high-concentration analysis. Conventional quartz-enhanced photoacoustic spectroscopy (QEPAS) typically employs acoustic micro-resonators (AmRs) coupled to quartz tuning forks (QTFs) to enhance trace-gas sensitivity. However, we reveal that at extremely high gas concentrations, concentration-dependent variations in the speed of sound detune the optimal acoustic matching condition of the AmR, causing severe signal distortion and restricting the dynamic range of conventional AmR-coupled QEPAS. To overcome this limitation, we develop a highly integrated dual-QTF QEPAS architecture combining an AmR-coupled QTF for sensitive trace detection and a bare QTF for robust linear measurements at high concentrations. The sensor enables continuous CO2 quantification from a minimum detection limit of 15 ppm to a 100% concentration, corresponding to a dynamic range of nearly five orders of magnitude. Dynamic measurements further demonstrate rapid gas exchange and real-time tracking of abrupt concentration variations. This work provides a compact and robust strategy for full-range gas monitoring in complex industrial environments.