Why the study?
Does remote monitoring reduce the number of inappropriate shocks related to lead fractures in patients with ICDs?
Does remote monitoring reduce the number of inappropriate shocks related to lead fractures in patients with ICDs?
Remote monitoring significantly reduces the burden of inappropriate shocks in ICD patients experiencing right ventricular lead fractures compared to conventional follow-up.
Should not yet change ICD follow-up practice; hypothesis-generating for remote monitoring in lead fracture cases.
AIMS: Lead fractures in implantable cardioverter-defibrillator (ICD) patients may cause inappropriate shocks (ISs). An early diagnosis is essential to prevent adverse clinical events. Implantable cardioverter-defibrillator remote monitoring (RM) permits prompt detection of lead fracture. Limited data define the impact of RM on ISs specifically related to lead fracture. We sought to compare the number of ISs related to lead fracture in patients with vs. without RM follow-up. METHODS AND RESULTS: We checked the registry of our institution and collected, between July 2007 and June 2014, 115 cases of right ventricular lead fractures. All relevant data were documented from patients' files, device-interrogation printouts and electronic records, and remote transmissions databases when applicable. We assessed the ISs that were related to lead fracture. The first study endpoint was the number of ISs per shocked patient. Among the 82 patients with conventional follow-up (CFU) and the 33 patients with RM, a first IS occurred to 32.9% (n = 27) and 30.3% (n = 10, P = 0.83) of the patients, respectively. Shocked patients in the RM group underwent significantly fewer ISs with a mean of 6 ± 2 shocks per patient [median of 3.5 shocks (2-8)] than those in the CFU group with a mean of 18 ± 5 shocks per patient [median of 10 shocks (5-22), P = 0.03]. CONCLUSION: Remote monitoring helps to reduce the burden of ISs related to ICD lead fractures.
No takes yet. Share an insight, caveat, or question.
Souissi et al. (2015) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: