Key result
Combined ultrasound vector flow parameters detect severe carotid stenosis with ~0.90 AUROC accuracy.
Why the study?
This study explored the feasibility and efficacy of using high-frame-rate ultrasound vector flow imaging to quantitatively assess hemodynamics in atherosclerotic internal carotid artery stenosis using turbulence index and wall shear stress.
Does combined quantitative assessment of turbulence index and wall shear stress using high-frame-rate ultrasound vector flow imaging improve diagnostic efficacy for evaluating moderate-to-severe internal carotid artery stenosis compared to single parameters?
Cross-Sectional (n=54)
No
Does combined quantitative assessment of turbulence index and wall shear stress using high-frame-rate ultrasound vector flow imaging improve diagnostic efficacy for evaluating moderate-to-severe internal carotid artery stenosis compared to single parameters?
Combining ultrasound vector flow imaging parameters of turbulence index and wall shear stress provides superior noninvasive diagnostic accuracy (AUROC 0.899) for distinguishing severe internal carotid artery stenosis compared to either hemodynamic metric alone.
Answers are grounded in this paper and cite their sources.
May enhance noninvasive carotid stenosis assessment; extends single-parameter data but leaves open prospective validation.
Objective This study explored the feasibility of using high‐frame‐rate ultrasound vector flow imaging (VFI) to quantitatively assess hemodynamics in atherosclerotic internal carotid artery stenosis (ICAS) by evaluating dual‐parameters, turbulence index (Tur), and wall shear stress (WSS). Their efficacy in evaluating carotid artery stenosis was also analyzed. Methods Fifty‐nine patients with ICAS were enrolled. B‐mode ultrasound and V Flow (a high‐frame‐rate VFI) were performed using the Resona R9 system. The stenosis rate was measured in grayscale mode, whereas the time‐averaged Tur index, the time‐averaged WSS (WSS mean ), and maximum WSS (WSS max ) around stenosis were measured. The combined diagnostic efficacy of Tur inand WSS was also investigated. Results Compared to proximal to stenosis (Tur index, 2.88% ± 3.65%), highly disordered blood flow was observed in the stenotic (23.17% ± 15.52%, p < 0.001) and distal segment (25.86% ± 17.29%, p < 0.001). WSS max ([11.91 ± 6.73] vs. [4.43 ± 5.4] Pa, p < 0.001) and WSS mean ([3.42 ± 2.67] vs. [0.86 ± 1.21] Pa, p < 0.001) were significantly bigger in stenotic than those in the distal segment. The differences in the ratio WSS max /Tur or WSS mean /Tur among different segments around stenosis were statistically significant ( p < 0.001). The combination of Tur index and WSS had the best diagnostic performance in ICAS (AUC, 0.899). Conclusion The application of Tur index and WSS for quantitative assessment of ICAS hemodynamic changes is feasible, with the combined evaluation of these two parameters providing incremental diagnostic value for carotid artery stenosis. VFI‐based dual quantitative parameters may offer promising noninvasive diagnostic tools for carotid artery stenosis in high‐stroke‐risk populations.
No takes yet. Share an insight, caveat, or question.
Hong et al. (2024) conducted a cross-sectional in Moderate-to-severe atherosclerotic internal carotid artery stenosis (ICAS) (n=54). High-frame-rate ultrasound vector flow imaging (VFI) dual-parameters: turbulence index (Tur) and wall shear stress (WSS) vs. Single-parameter assessment (Tur index alone or WSS alone) and prestenosis/poststenosis vessel segments was evaluated on Diagnostic performance (AUROC) of combined Tur index and WSS in distinguishing severe internal carotid artery stenosis (p=<0.001). Combined ultrasound vector flow imaging parameters (turbulence index and wall shear stress) achieved an AUROC of 0.899 for diagnosing severe carotid stenosis.
Synapse has enriched 10 related papers. Consider them for comparative context: