Why the study?
Does a rapid nonmotorized treadmill exercise protocol identify suboptimal rate responsiveness in patients with rate responsive pacemakers compared to normal physiological responses?
Does a rapid nonmotorized treadmill exercise protocol identify suboptimal rate responsiveness in patients with rate responsive pacemakers compared to normal physiological responses?
A rapid (<5 minute) nonmotorized treadmill test demonstrates that nominal factory settings for rate responsive pacemakers often fail to mimic physiological heart rate increases during exertion.
Answers are grounded in this paper and cite their sources.
Nominal rate-responsive pacemaker settings may underdeliver chronotropic response; this Level 4 observation is hypothesis-generating and requires prospective validation.
BACKGROUND: Rate responsive (RR) pacemakers are commonly implanted with nominal conservative factory-set responsiveness, which is usually accepted because established exercise protocols are time-consuming. We aimed for efficient assessment of RR pacemaker settings. METHODS: We tested exercise heart rates in controls and paced patients using a brief exercise test that approximates real-life levels of exertion. The test used a nonmotorized treadmill: 30 seconds walking at patient-determined speed followed by 15 seconds brisk exertion. Subjects totaled 110: 26 with RR pacemakers; 22 with non-RR pacers; 27 "sick" nonpaced control patients; and 35 healthy controls. Heart rate (HR) was measured prior to exercise, after 30 seconds of casual walk, after 15 seconds of brisk walk, and 1 minute into recovery. Testing required <5 minutes from set-up to recovery. RESULTS: The 26 RR pacer patients had a mean HR at rest = 74 +/- 10 beats per minute (bpm), walk = 87 +/- 14, and brisk = 94 +/- 18 (increase 27%). Non-RR pacer patients (n = 22): rest = 73 +/- 12 bpm, walk = 88 +/- 14, and brisk = 94 +/- 17 (increase 24.3%, P = 0.60 vs RR patients). "Sick" controls (n = 27): rest = 78 +/- 14 bpm, walk = 102 +/- 17, and brisk = 117 +/- 18 (increase 51.9%, P< 0.001 vs RR pts). For the healthy controls, HRs were at rest 83+/11 bpm, walk = 104 +/- 12, and brisk = 117 +/- 13 (P< 0.001 compared to both paced groups; P = NS vs sick controls). CONCLUSIONS: Nominal RR settings may be suboptimal for many patients. The nonmotorized treadmill test allows quick and inexpensive assessment of RR programming, with the potential for efficient RR optimization.
No takes yet. Share an insight, caveat, or question.
SHABER et al. (2008) studied this question.