Why the study?
The study was conducted to evaluate the utility of pulse oximetry as a diagnostic test to detect peripheral arterial disease compared with arterial Doppler ultrasonography, and to assess ABPI efficacy.
Does pulse oximetry accurately detect peripheral arterial disease in patients aged 40 years or older with symptoms or risk factors for PAD?
Does pulse oximetry accurately detect peripheral arterial disease in patients aged 40 years or older with symptoms or risk factors for PAD?
Pulse oximetry offers a simple, accessible tool with moderate sensitivity and specificity for detecting peripheral arterial disease.
Pulse oximetry may not reliably detect PAD in symptomatic or at-risk patients; leaves open the need for prospective validation before any clinical adoption.
A BSTRACT Use of pulse oximeter as a tool to detect peripheral arterial disease. A cross-sectional study in a tertiary care center. Aims and Objectives: To evaluate the utility of pulse oximeter (index test) as a diagnostic test to detect peripheral arterial disease (PAD) by comparing it with the standard diagnostic test, that is, arterial Doppler ultrasonography. Secondary Objective: To evaluate the efficacy of ankle brachial pressure index (ABPI) in detecting PAD. Materials and Methods: Patients of age 40 years or older with symptoms of PAD or asymptomatic patients with risk factors for peripheral vascular disease had pulse oximeter readings of upper and lower limbs. A decrease in pulse oximeter reading of 2% in the lower limb compared to the upper limb reading was considered positive. The patients then underwent ABPI measurement with a handheld Doppler device, arterial Doppler ultrasonography of the lower limb was conducted, and the results of each test were compared. Results: The values obtained by pulse oximeter measurement had a sensitivity of 64.4% (95% CI 56.7%-71.4%), specificity of 75% (95% CI 47.6%-92.7%), PPV and NPV of 96.6% (95% CI 92.2%-98.5%) and 16.2% (95% CI 12.04%-21.48%) respectively, with a P value of 0.002 and accuracy of 65.2%, whereas values obtained by ABPI had a sensitivity of 72.4% (95% CI 65.5%-79.1%), specificity of 50% (95% CI 24.6%-75.3%), PPV and NPV of 94% (95% CI 90.6%-96.3%) and 14.3% (95% CI 8.8%-22.3%) respectively, with a P value 0.06, and an accuracy of 70%. Conclusion: Pulse oximeter can be used as a tool for the detection of PAD with considerable accuracy.
No takes yet. Share an insight, caveat, or question.
Acharya et al. (2025) studied this question.