Why the study?
Established studies focused on myocardial effects during galvanic coupling conductive intracardiac communication for leadless pacemakers, but the impact of blood volume variations during the cardiac cycle was not well understood.
Blood volume variations during the cardiac cycle significantly impact the signal transmission gain of intracardiac communication channels, particularly the RV-RA channel, which is crucial for optimizing leadless pacemaker systems.
Blood volume effects on intracardiac gain merit design consideration for leadless systems; animal data leaves open human translation.
In galvanic coupling conductive intracardiac communication(GCCIC) of the leadless pacemakers, the electrical signal transmitted directly through the myocardium and blood is inevitably affected by the cardiac cycle. Established studies focused more on the effect of the myocardium. However, our preliminary in-vitro experiments suggested that blood volume variations also significantly impacted signal transmission. In this article, we analyzed the blood volume variations during the cardiac cycle and designed an in-vitro experimental platform containing a simulated heart beating system and an automatic channel characteristic acquisition system, which controlled two peristaltic pumps to realize the periodic blood volume variations and the continuous acquisition of channel gain. Through the in-vitro porcine heart experiment, the effect of frequency and blood volume variations during the cardiac cycle on two channel gains was analyzed. Considering the impact of high-frequency signal leakage, the channel gain variations of the low frequency are the main concern. The results showed that the channel gain was positively correlated with frequency; it changed periodically with blood volume variations in the cardiac cycle, and the trends were different due to the different signal paths of the two channels. For the Right Ventricle-Right Atrium channel, the gain varied from$-67$dB to$-53$dB and is inversely correlated with blood volume. The gain fluctuation range was smaller for the Right Ventricle-Left Ventricle channel, about 2 dB. This study shows that the gain of intracardiac communication channels, especially the RV-RA channel, is influenced by blood volume variations during the cardiac cycle.
No takes yet. Share an insight, caveat, or question.
Chen et al. (2022) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: