Answers are grounded in this paper and cite their sources.
An ultra-highly sensitive and robust CH 4 sensor is reported based on a 3.3 µm interband cascade laser (ICL) and a low-noise differential photoacoustic (PAS) cell. The ICL emission wavelength targeted a fundamental absorption line of CH 4 at 2988.795 cm −1 with an intensity of 1.08 × 10 −19 cm/molecule. The double-pass and differential design of the PAS cell effectively enhanced the PAS signal amplitude and decreased its background noise. The wavelength modulation depth, operating pressure and V-T relaxation promotion were optimized to maximize the sensor detection limit. With an integration time of 90 s, a detection limit of 0.6 ppb was achieved. No additional water or air laser cooling were required and thereby allowing the realization of a compact and robust CH 4 sensor.
No takes yet. Share an insight, caveat, or question.
Zheng et al. (2020) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: