Why the study?
How can true atrial undersensing be differentiated from normal pacemaker operation in patients with DDD pacemakers showing dropped P waves?
How can true atrial undersensing be differentiated from normal pacemaker operation in patients with DDD pacemakers showing dropped P waves?
Answers are grounded in this paper and cite their sources.
Careful electrocardiographic evaluation can differentiate true atrial undersensing from normal pacemaker operation, avoiding unnecessary reinterventions.
ECG tracings of three patients in whom AV universal (DDD) pacemakers were implanted intermittently demonstrated dropped P waves. In one patient, true atrial undersensing was present; in the others, sensing of the atrial electrode was appropriate, but sensing of sinus P waves was intermittently blocked by normal pacemaker operation. In this report we discuss the electrocardiographic diagnosis of atrial undersensing in order to avoid unnecessary reinterventions.
No takes yet. Share an insight, caveat, or question.
Mechelen et al. (1986) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: