Key result
Decreasing ventricular sensitivity resolved intermittent P wave sensing caused by T wave oversensing by the ventricular lead in a patient with a DDD pacemaker.
Decreasing ventricular sensitivity can resolve intermittent P wave sensing caused by T wave oversensing in patients with DDD pacemakers.
Answers are grounded in this paper and cite their sources.
May guide oversensing troubleshooting in DDD pacemakers; hypothesis-generating from one case and should not yet change practice.
We present the case of a patient with a DDD pacemaker and intermittent P wave sensing due to T wave oversensing by the ventricular lead. The T wave sensing caused initiation of an extended atrial refractory period and the P waves, falling within this period, were not sensed. The problem was solved by decreasing the ventricular sensitivity.
No takes yet. Share an insight, caveat, or question.
Paraskevaidis et al. (1999) studied this question. Decreasing ventricular sensitivity resolved intermittent P wave sensing caused by T wave oversensing by the ventricular lead in a patient with a DDD pacemaker.