Key result
Decreasing ventricular sensitivity resolved intermittent P wave sensing caused by T wave oversensing by the ventricular lead in a patient with a DDD pacemaker.
Population
A patient with a DDD pacemaker experiencing intermittent P wave sensing due to T wave oversensing by the…
Design
Case_report
Authors
Loading...
May guide oversensing troubleshooting in DDD pacemakers; hypothesis-generating from one case and should not yet change practice.
Decreasing ventricular sensitivity can resolve intermittent P wave sensing caused by T wave oversensing in patients with DDD pacemakers.
Paraskevaidis et al. (1999) studied this question. Decreasing ventricular sensitivity resolved intermittent P wave sensing caused by T wave oversensing by the ventricular lead in a patient with a DDD pacemaker.