Key result
Cardiac function index correlates moderately with echographic LVEF and tracks therapy-induced changes.
Why the study?
Whether the cardiac function index provided by transpulmonary thermodilution behaves as a reliable indicator of left ventricular systolic function during fluid loading and inotropic stimulation was untested against echographic left ventricular ejection fraction.
Does the cardiac function index provided by transpulmonary thermodilution accurately reflect left ventricular systolic function and track changes during therapeutic interventions in patients with acute circulatory failure?
Observational (n=39)
No
Does the cardiac function index provided by transpulmonary thermodilution accurately reflect left ventricular systolic function and track changes during therapeutic interventions in patients with acute circulatory failure?
The cardiac function index derived from transpulmonary thermodilution is a reliable bedside indicator of left ventricular global systolic function that accurately tracks the effects of inotropic therapy.
Answers are grounded in this paper and cite their sources.
May support cardiac function index as LVEF surrogate in acute failure; hypothesis-generating, needs randomized validation.
Objective To test whether cardiac function index could actually behave as an indicator of left ventricular systolic function by testing if 1) it increased with inotropic stimulation; 2) it was not altered by fluid loading; 3) it correlated with the echographic left ventricular ejection fraction, considered as the clinical gold standard for measuring left ventricular systolic function, and it reliably tracked the changes in left ventricular ejection fraction during therapeutic intervention. The transpulmonary thermodilution calculates the cardiac function index, which is the ratio of cardiac output over global end-diastolic volume. Design Prospective study. Setting Medical intensive care unit of a university hospital. Patients Thirty-nine patients (n = 48 cases) with acute circulatory failure. Intervention A 500-mL saline administration (n = 24 cases) and a dobutamine infusion (n = 24 cases). Measurements and main results We simultaneously measured left ventricular ejection fraction (monoplane or biplane Simpson method) and cardiac function index at baseline and after saline and dobutamine administration. As volume expansion altered neither left ventricular ejection fraction (47 +/- 11% to 47 +/- 11%) nor cardiac function index (4.5 +/- 2.2 to 4.5 +/- 2.1 min(-1)), dobutamine infusion significantly increased left ventricular ejection fraction by 32 +/- 28% and cardiac function index by 29 +/- 22%. Considering the 96 cardiac function index:left ventricular ejection fraction pairs of measurements, cardiac function index and left ventricular ejection fraction were correlated significantly (r = .67, p < .05). A cardiac function index <3.2 min(-1) predicted a left ventricular ejection fraction of <35% with a sensitivity of 81% and a specificity of 88%. Importantly, the changes in left ventricular ejection fraction and in cardiac function index during fluid and dobutamine administration were correlated significantly (r = .79, p < .05). Conclusions Cardiac function index fulfilled the criteria required from a clinical indicator of left ventricular global systolic function. Thus, it could serve as an easy bedside detection of the alteration in left ventricular ejection fraction that should alert the physician and incite to perform an echocardiography. Furthermore, cardiac function index tracked accurately the effects of an inotropic therapy.
No takes yet. Share an insight, caveat, or question.
Jabot et al. (2009) conducted an observational in Acute circulatory failure (n=39). Cardiac function index via transpulmonary thermodilution (with 500-mL saline and dobutamine infusion) vs. Echographic left ventricular ejection fraction (Simpson method) was evaluated on Correlation between cardiac function index and echographic left ventricular ejection fraction (r = 0.67, p=<0.05). Cardiac function index correlated with echographic left ventricular ejection fraction and tracked therapy-induced changes.