Why the study?
Late lead perforation is a rare diagnosis with little data existing regarding management practices and outcomes.
Is transvenous management safe and effective for clinically significant late lead perforation in patients with cardiac implantable electronic devices?
Observational (n=11)
No
Is transvenous management safe and effective for clinically significant late lead perforation in patients with cardiac implantable electronic devices?
Late lead perforation can often be safely managed with a cautious transvenous approach, provided intraprocedural ultrasound and immediate surgical backup are available.
Answers are grounded in this paper and cite their sources.
Supports cautious transvenous extraction with backup in select late perforations; leaves open need for confirmatory prospective data.
INTRODUCTION: Late lead perforation (LLP), defined as perforation ≥30 days from cardiac implantable electronic device implant, is a rare diagnosis and little data exist regarding management practices and outcomes. The purpose of this study was to evaluate the occurrence, safety, and efficacy of transvenous management of clinically significant LLP. METHODS: The electronic medical records of a single-center tertiary hospital were reviewed for all patients who were referred for LLP or its sequelae. RESULTS: Eleven consecutive patients were identified from October 2011 to December 2018 with clinically significant LLP. Patients most often presented with pericardial symptoms with the exception of one asymptomatic patient. The median time from lead implant to intervention for LLP was 246 days. Nine patients were managed with an initial transvenous approach, with one requiring sternotomy (lead 6.3 years old). Two patients had a surgical approach, one performed at an outside hospital with subsequent death and another had a mini-thoracotomy, but the lead was removed percutaneously with no surgical repair. In this small cohort, there was no association between the lead extending beyond the parietal pericardium and surgical repair (P = .99). CONCLUSION: Our single-center experience suggests that LLP can be initially managed with a cautious transvenous approach in most patients, but intraprocedural ultrasound for pericardial monitoring and a rescue plan with immediate surgical back up is mandatory.
No takes yet. Share an insight, caveat, or question.
Jessel et al. (2020) conducted an observational in Late lead perforation (LLP) (n=11). Transvenous management was evaluated on Occurrence, safety, and efficacy of transvenous management of clinically significant LLP.