Key result
Implantation of a dual-chamber pacemaker was complicated by simultaneous bilateral pneumothorax due to left subclavian vein puncture and right atrial electrode perforation.
Why the study?
Pneumothorax usually arises from needle injury during subclavian vein access and rarely from electrode perforation, with no prior reports of simultaneous bilateral pneumothorax from both mechanisms.
Case Report (n=1)
This first reported case of simultaneous bilateral pneumothorax during pacemaker implantation highlights the need for careful venous access and lead positioning in high-risk patients.
Answers are grounded in this paper and cite their sources.
Rare bilateral pneumothorax in pacemaker implantation warrants procedural vigilance; leaves open optimal prevention strategies for high-risk patients.
Pneumothorax occurs mostly due to needle injury of the pleura when trying to get access to the subclavian vein and rarely due to electrode perforation. The present case report is the first case presented about acute simultaneous iatrogenic bilateral pneumothorax due to puncture of the left subclavian vein and electrode perforation of the atrial wall, the pericardium, and the pleura. Risk factors, and how to avoid these complications, are highlighted, and symptoms, diagnostics, and management of pneumothorax and cardiac perforation are described.
No takes yet. Share an insight, caveat, or question.
Line Lisbeth Olesen (2020) conducted a case report in Third-degree atrioventricular block (n=1). Dual-chamber pacemaker implantation was evaluated on Bilateral pneumothorax. Implantation of a dual-chamber pacemaker was complicated by simultaneous bilateral pneumothorax due to left subclavian vein puncture and right atrial electrode perforation.
Synapse has enriched 10 related papers. Consider them for comparative context: