Why the study?
Does automatic blood pressure measurement by the Keito machine accurately reflect office blood pressure measured by physicians in patients scheduled for pre-catheterisation screening?
Does automatic blood pressure measurement by the Keito machine accurately reflect office blood pressure measured by physicians in patients scheduled for pre-catheterisation screening?
The Keito machine significantly underestimates blood pressure compared to physician office measurements and is not recommended for medical screening or follow-up of hypertension.
Answers are grounded in this paper and cite their sources.
Keito readings may underestimate office BP in pre-catheterization patients; leaves open accuracy for hypertension screening pending validation studies.
BACKGROUND: The Keito machine offers automatic measurements of blood pressure (BP), height and weight on insertion of coins and has been introduced in pharmacies. DESIGN: Cross-sectional study comparing automatic BP measurements by the Keito machine to office BP measurements by physicians. METHODS: Patients scheduled for pre-catheterisation screening participated in the study. Their BP was first measured using the Keito machine, then by physicians. Office BP was recorded as the last of three consecutive BP measurements recorded with one-min intervals after a five-min rest in the sitting position. In a sub-study BP was measured simultaneously during the Keito measurement by a physician. RESULTS: In 390 consecutive patients average BP was significantly lower with the Keito machine compared to office BP measurements made by the physicians (136/75+/-19/8 mmHg versus 141/79+/-21/10 mmHg, both p<0.001). The correlation coefficient (r) was 0.56 (p<0.001) for systolic BP (SBP) and 0.53 (p<0.001) for diastolic BP (DBP). Bland-Altman analysis showed a mean difference (+/-2 SD) for SBP and DBP of -5 (+/-37) and -4 (+/-17) mmHg, respectively. When defining hypertension (HT) as office SBP> or =140 and/or DBP> or =90 mmHg, the Keito method diagnosed 83% of the systolic and 62% of the diastolic hypertensive population correctly. The classification of systolic and diastolic normotensive was correct in 61% and 86%, respectively. CONCLUSION: Agreement between office and Keito BP is poor. The Keito machine underestimates SBP on average by 5 mmHg and DBP by 4 mmHg, which may be of significance for diagnosing HT and starting anti-hypertensive therapy. However, the difference can be much larger in individual patients. Therefore, the Keito machine cannot be recommended for medical screening of HT or as a replacement for follow-up by physicians.
No takes yet. Share an insight, caveat, or question.
Staal et al. (2004) studied this question.