Answers are grounded in this paper and cite their sources.
A portable laser induced breakdown spectrometer (LIBS), built around a microscope using a Q-switched Nd:GGG laser releasing multiple pulses and a non-intensified fiber optic spectrometer, was used to investigate the signal improvement of emission data achievable by the multiple-pulse approach. LIBS experiments were carried out in air and using Al, Cu, Si and Zn metallic targets. It was found that the use of 7 laser pulses, separated by ca. 25 μs time intervals, enhance line emissions by a factor of 2.5 to 129 with respect to the single-pulse case, and also decrease the relative standard deviation of the signal based on five consecutive measurements to 2–5%. Time-resolved emission data and the observation of ablation craters suggest that the signal enhancement is mainly attributable to the reheating of the plasma by multiple laser pulses.
No takes yet. Share an insight, caveat, or question.
Galbács et al. (2005) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: