Why the study?
To analyze auto-annotated radiofrequency ablation data from a pulmonary vein isolation procedure that used widely accepted methods but resulted in an atrioesophageal fistula, aiming to improve safety guidelines.
Demonstrates that atrioesophageal fistula can occur during PVI even when adhering to widely accepted RF ablation parameters.
Even standard RF parameters leave atrioesophageal fistula risk uneliminated; leaves open need for better prevention strategies.
We present detailed analyses of auto-annotated radiofrequency (RF) ablation data from a pulmonary vein isolation (PVI) procedure that was performed using widely accepted methods (30 W, ≤6 mm inter–ablation site distance, ablation index guidance) but still ended with a catastrophic complication. We hope that lessons learned in this case will help inform clinical practice guidelines toward greater safety during atrial fibrillation (AF) ablation.