Why the study?
DOACs interfere in laboratory coagulation testing, prompting an evaluation of whether commercial DOAC removal methods eliminate DOAC concentrations and false positive results in lupus anticoagulant testing.
Does ex vivo DOAC removal reduce false positive results in lupus anticoagulant testing in patient samples with high DOAC concentrations?
Does ex vivo DOAC removal reduce false positive results in lupus anticoagulant testing in patient samples with high DOAC concentrations?
Ex vivo DOAC removal using commercial methods effectively reduces DOAC concentrations and halves the rate of false positive lupus anticoagulant screening tests, reducing the need for unnecessary repeated testing.
May reduce false-positive LAC screening; leaves open validation before routine lab adoption.
Direct oral anticoagulants (DOAC) interfere in laboratory coagulation testing. The aim here was to study how commercial DOAC removal methods, DOAC Filter® and DOAC-Stop™, perform to eliminate DOAC concentrations and false positive results in lupus anticoagulant (LAC) testing. We acquired 50 patient samples with high concentrations of DOACs: apixaban (n = 18, range 68–572 ng/mL), dabigatran (n = 8, range 47–154 ng/mL), edoxaban (n = 8, range 35–580 ng/mL) and rivaroxaban (n = 16, range 69–285 ng/mL). DOACs were removed ex vivo with either DOAC Filter® (n = 28) or DOAC-Stop™ (n = 22). Additionally, commercial control and calibrator samples were studied (n = 13 for DOAC Filter®, n = 14 for DOAC-Stop™). LAC screening was performed before and after DOAC removal. Both DOAC Filter® and DOAC-Stop™ were effective in removing DOAC concentrations in samples: DOAC concentrations decreased to median of 0 ng/mL (range 0–48 ng/mL). Only one sample had more than residual 25 ng/mL of DOAC (apixaban). Before DOAC removal, 96% (48/50) of patient samples and over 90% (12/13 DOAC Filter®, 13/14 DOAC-Stop™) of control/calibrator samples were positive in the LAC screening. In patient samples, LAC screening turned negative in 61% (17/28) after DOAC Filter® and 45% (10/22) after DOAC-Stop™ treatment. All control samples became negative after DOAC removal. In conclusion, DOAC removal ex vivo reduces false positives in LAC screening. DOAC removal halved the need for confirmation or mixing tests- Although a subset of patients would require further testing, DOAC removal reduces unnecessary repeated LAC testing.
No takes yet. Share an insight, caveat, or question.
Savola et al. (2022) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: