Patients with metal spinal implants often undergo MRI, but the metal distorts the magnetic field causing artifacts that obscure the surrounding tissue. The susceptibility effect is proportional to field strength, and is therefore more pronounced at 1.5T and 3T compared to lower field strengths. We compared a contemporary mid-field MRI (0.55T) with standard 3T MRI in ten people with spinal hardware. For patients with large lower-back implants, 0.55T showed the surrounding tissues more clearly and gave radiologists greater confidence, even though images looked slightly noisier. Specialized artifact-reduction sequences added little benefit in this small group.