Key result
ICD reprogramming with extended AV delay eliminates recurrent polymorphic VT/VF from R-on-T pacing.
Why the study?
ICDs and biventricular ICDs can be proarrhythmic, particularly when pacing into scar tissue.
Does reprogramming the atrioventricular delay and lower rate limit prevent recurrent polymorphic VT/VF caused by masked PVCs during the post-atrial pacing ventricular blanking period in patients with dual-chamber ICDs?
Case Report (n=2)
No
Does reprogramming the atrioventricular delay and lower rate limit prevent recurrent polymorphic VT/VF caused by masked PVCs during the post-atrial pacing ventricular blanking period in patients with dual-chamber ICDs?
This case report identifies a novel proarrhythmic mechanism in dual-chamber ICDs with normal lead parameters—where PVCs masked in the post-atrial pacing ventricular blanking period during pseudo-pseudofusion trigger R-on-T pacing and VF—which can be completely resolved by reprogramming the AV delay and lower rate limit.
Answers are grounded in this paper and cite their sources.
Reprogramming may eliminate masked PVC-triggered VT/VF in select dual-chamber ICD patients; leaves open generalizability beyond these cases.
Implantable cardioverter-defibrillators (ICD) are indicated for patients with cardiomyopathies to reduce their risk of sudden cardiac death from both ventricular tachycardia (VT) and ventricular fibrillation (VF). ICDs and biventricular ICDs can, however, be proarrhythmic, especially when pacing into a scar. We describe 2 cases of patients with frequent premature ventricular contractions (PVCs) who developed recurrent polymorphic VT/VF due to interaction between PVC timing and blanking periods, leading to R-on-T ventricular pacing following pseudo-pseudofusion.
No takes yet. Share an insight, caveat, or question.
H. et al. (2020) conducted a case report in Cardiomyopathy with dual-chamber implantable cardioverter-defibrillator (ICD), frequent premature ventricular contractions (PVCs), and recurrent polymorphic ventricular tachycardia (VT) or ventricular fibrillation (VF) (n=2). ICD reprogramming (extending atrioventricular delay to 400 ms and adjusting lower rate limit to 40 bpm or 70 bpm) was evaluated on Recurrence of polymorphic ventricular tachycardia (VT) or ventricular fibrillation (VF) episodes. ICD reprogramming by extending the AV delay to 400 ms and adjusting the lower rate limit eliminated recurrent polymorphic VT/VF caused by masked PVCs and R-on-T pacing in all 2 patients.