Answers are grounded in this paper and cite their sources.
Superparamagnetic labels, 400 nm dextran iron oxide particles and 2 μm polymer encapsulated iron oxide microspheres, with biomolecules immobilized on the surface, e.g., the enzyme horseradish peroxidase (20–40 molecules per label) were controllably placed on chip sites (5×15 μm2) using tapered Al current lines (10–20 mA current) and moved to and from adjacent spin valve sensors [2×6 μm,2, magnetoresistance (MR) ∼5%]. Average MR signals of 1.2 and 0.6 mV were obtained for the detection of bulk numbers of 400 nm and 2 μm labels respectively using an on-chip field of 15 Oe and a sense current of 5 mA. The moment per label was calculated at 5×10−13 emu for the 400 nm labels and 5×10−12 emu for the 2 μm labels, illustrating the higher density of the 400 nm particles. MR signals of ∼100 μV were obtained for single 2 μm labels positioned over the spin valve sensor using an on-chip field of 15 Oe and 8 mA sense current. The corresponding sensor saturation occurred at ∼1 mV, with a noise level of ∼10 μV. The estimated maximum MR signal for one 2 μm label directly on top of the sensor was ∼400 μV. Biotechnological applications include high sensitivity biosensors and biochips for protein and DNA screening.
No takes yet. Share an insight, caveat, or question.
Graham et al. (2002) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: