Why the study?
The impact of transcatheter prosthesis frame design features on implantation outcomes in the aortic root model required analysis to improve functional characteristics.
Answers are grounded in this paper and cite their sources.
This article presents an analysis of the impact of the transcatheter prosthesis frame design features on the results of its implantation in the aortic root model. In this paper we analyzed the various approaches to the design of such structures, as well as modifications in order to improve their functional characteristics during the implantation. As a general method for obtaining the results of interaction of the objects was used finite element method with nonlinear materials description and analysis of the main parameters: the stress-strain state, radial and friction forces.
No takes yet. Share an insight, caveat, or question.
Овчаренко et al. (2015) studied Transcatheter aortic valve prosthesis implantation (In Silico Model). Modified transcatheter prosthesis frame designs (specifically increased distal zone diameter) vs. Classic universal tubular stent design was evaluated on Friction force, radial force, and mechanical stress (von Mises) during simulated implantation.