Key result
A transcatheter aortic valve frame design with 12 cells, 0.3 mm strut width, and 0.4-0.5 mm thickness provides the optimal balance of radial and pinching forces while minimizing strain and stress.
Answers are grounded in this paper and cite their sources.
Finite element analysis of TAVI frame cell geometry parameters can assist in the development of new artificial heart valves and evaluation of existing prostheses.
This paper presents the analysis of dependences between frame basic cell geometry parameters and function via finite element analysis. The simplified models of frame cell with varied strut width, thickness and quantity in a circle was researched to evaluate radial forces, maximum stress and strain, permanent residual strain and pinching load forces. The outcomes of this study might help in the development of new artificial heart valves and during the analysis of existing in-clinical TAVI prostheses.
No takes yet. Share an insight, caveat, or question.
Овчаренко et al. (2014) studied Aortic valve disease. TAVI frame cell geometry optimization was evaluated on Radial forces, stress, strain, and pinching forces. A transcatheter aortic valve frame design with 12 cells, 0.3 mm strut width, and 0.4-0.5 mm thickness provides the optimal balance of radial and pinching forces while minimizing strain and stress.
Synapse has enriched 10 related papers. Consider them for comparative context: