Why the study?
Does a stepwise mapping approach optimize the real-time assessment of pulmonary vein disconnection during cryoablation of atrial fibrillation?
Does a stepwise mapping approach optimize the real-time assessment of pulmonary vein disconnection during cryoablation of atrial fibrillation?
A systematic stepwise mapping approach using an inner circular mapping catheter allows for real-time monitoring of pulmonary vein disconnection in nearly all cases during AF cryoablation.
Answers are grounded in this paper and cite their sources.
May support stepwise mapping for real-time PV monitoring in cryoablation; leaves open impact on outcomes in prospective trials.
AIM: Real-time assessment of pulmonary vein (PV) disconnection is possible using an inner circular mapping catheter during cryoablation of atrial fibrillation (AF). However, it has been recently demonstrated that such continuous monitoring may only be possible in <50% of PVs. We hypothesized that a stepwise mapping approach, including pacing manoeuvres, could optimize monitoring of real-time PV disconnection during ablation. METHODS AND RESULTS: Single-centre, prospective observational study (NCT01843465) including 34 consecutive eligible patients (128 PVs) undergoing a first procedure of cryoballoon ablation of AF using the Artic Front Advance(®) 28 mm catheter and a 20 mm diameter Achieve(®) catheter (AC) in all cases. Monitoring of real-time entrance block was possible, when AC was maintained in the standard position (distal to the tip of the Artic Front Advance(®) catheter) in 47 (36.7%) PVs. In an additional 63 cases (49.2%), such monitoring was possible after moving AC to a more proximal position and using different torqueing movements. Finally, using supplemental systematic pacing manoeuvres to test exit block, real-time assessment of PV disconnection was possible in 15 (11.7%) more PVs. Overall, real-time assessment of PV disconnection was possible in 97.7% of cases, after a mean duration of 48.6 ± 33.0 s. CONCLUSION: Our results suggest that optimal use of the AC, with a systematic stepwise mapping approach, may dramatically improve the real-time monitoring of PV disconnection during AF cryoablation.
No takes yet. Share an insight, caveat, or question.
Bovéda et al. (2013) studied this question.