Key result
The Achieve mapping catheter identified pulmonary vein reconnections with a positive predictive value of 100% and negative predictive value of 96% compared to ICE-guided mapping.
Why the study?
Substrate alterations in patients with AF recurrence after using the Arctic Front Cryoballoon System have not been well characterized.
Does the Achieve circular mapping catheter accurately detect pulmonary vein reconnections compared to ICE-guided mapping in patients undergoing redo AF ablation after cryoablation?
Observational (n=15)
No
Does the Achieve circular mapping catheter accurately detect pulmonary vein reconnections compared to ICE-guided mapping in patients undergoing redo AF ablation after cryoablation?
The Achieve circular mapping catheter accurately identifies pulmonary vein reconnections during redo AF ablation following index cryoballoon ablation.
BACKGROUND: The Arctic Front Cryoballoon System is a technology in which substrate alterations in patients with atrial fibrillation (AF) recurrence have not been well characterized. In this study, we evaluated sites of pulmonary vein (PV) reconnections and the accuracy of the Achieve™ circular mapping catheter in detecting these reconnections after cryoablation. METHODS: This study included 15 patients undergoing redo AF ablation after a prior single cryoablation procedure. PV reconnection sites were determined by measuring PV signals and high output pacing from 4 vectors of the Achieve catheter. The results were compared with a roving mapping catheter guided by rotational intracardiac echocardiography (ICE) in the left atrium. RESULTS: All patients had PV reconnections (2.1 ± 0.8 veins/patient). The left superior PV was most commonly reconnected (n = 11), whereas the right inferior PV was least likely (n = 3). Both carinas (left: n = 11; right: n = 7) and left atrial appendage ridge (n = 11) were also frequently reconnected. Mapping with the Achieve catheter showed a positive predictive value (PPV) 100% and negative predictive value (NPV) 96% when compared with ICE guided mapping. In 2 patients, right superior PV reconnection was not identified by the Achieve. CONCLUSION: During redo AF ablation after index cryoablation, multiple PVs are usually reconnected, with both carinas and left atrial appendage ridge being common sites of reconnection. The Achieve mapping catheter was able to identify reconnection with high positive and negative predictive values.
No takes yet. Share an insight, caveat, or question.
Shah et al. (2019) conducted an observational in Atrial Fibrillation (n=15). Achieve mapping catheter vs. Intracardiac echocardiography (ICE) guided mapping was evaluated on Detection of pulmonary vein reconnection (Positive Predictive Value). The Achieve mapping catheter identified pulmonary vein reconnections with a positive predictive value of 100% and negative predictive value of 96% compared to ICE-guided mapping.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: