Why the study?
MRI of excised hearts at ultra-high field strengths (B0 >= 7 T) can provide high-resolution, high-fidelity ground truth data for biomedical studies, imaging science, and artificial intelligence.
Does a custom-built 8Tx/16Rx pTx coil improve receive sensitivity and parallel imaging capability compared to a commercial 1Tx/32Rx head coil for 7 T MRI of excised porcine hearts?
Does a custom-built 8Tx/16Rx pTx coil improve receive sensitivity and parallel imaging capability compared to a commercial 1Tx/32Rx head coil for 7 T MRI of excised porcine hearts?
A dedicated 8Tx/16Rx transceiver array for 7 T MRI provides superior receive sensitivity and parallel imaging capabilities for ultra-high-resolution ex-vivo cardiac imaging compared to standard head coils.
Answers are grounded in this paper and cite their sources.
May improve ex-vivo 7T cardiac imaging quality; leaves open in-vivo validation and clinical translation.
Introduction MRI of excised hearts at ultra-high field strengths ( B₀ B 0 ≥7 T) can provide high-resolution, high-fidelity ground truth data for biomedical studies, imaging science, and artificial intelligence. In this study, we demonstrate the capabilities of a custom-built, multiple-element transceiver array customized for high-resolution imaging of excised hearts. Method A dedicated 16-element transceiver loop array was implemented for operation in parallel transmit (pTx) mode (8Tx/16Rx) of a clinical whole-body 7 T MRI system. The initial adjustment of the array was performed using full-wave 3D-electromagnetic simulation with subsequent final fine-tuning on the bench. Results We report the results of testing the implemented array in tissue-mimicking liquid phantoms and excised porcine hearts. The array demonstrated high efficiency of parallel transmits characteristics enabling efficient pTX-based B 1 + -shimming. Conclusion The receive sensitivity and parallel imaging capability of the dedicated coil were superior to that of a commercial 1Tx/32Rx head coil in both SNR and T 2 *-mapping. The array was successfully tested to acquire ultra-high-resolution (0.1 × 0.1 × 0.8 mm voxel) images of post-infarction scar tissue. High-resolution (isotropic 1.6 mm 3 voxel) diffusion tensor imaging-based tractography provided high-resolution information about normal myocardial fiber orientation.