Answers are grounded in this paper and cite their sources.
This study aims to develop an improved Feldkamp-Davis-Kress (FDK) reconstruction algorithm anisotropic total variation (ATV) minimization to enhance the image quality of low-dose conebeam tomography (CBCT). The algorithm first applies a filter that integrates the Shepp- filter into a cosine window function on all projections for impulse noise removal. A total objective function with anisotropic penalty is then minimized to enhance the difference the real structure and noise using the steepest gradient descent optimization with step sizes. The preserving parameter to adjust the separation between the noise-free and areas is determined by calculating the cumulative distribution function of the gradient of the filtered image obtained by the application of the filtering operation on each. With these minimized ATV projections, voxel-driven backprojection is finally performed generate the reconstructed images. The performance of the proposed algorithm was evaluated the catphan503 phantom dataset acquired with the use of a low-dose protocol. Qualitative quantitative analyses showed that the proposed ATV minimization provides enhanced CBCT images compared with those generated by the conventional FDK algorithm, with a contrast-to-noise ratio (CNR), lower root-mean-square-error, and higher correlation. The algorithm not only leads to a potential imaging dose reduction in repeated CBCT scans lower mA levels, but also elicits high CNR values by removing noisy corrupted areas and by the heavy penalization of striking features.
No takes yet. Share an insight, caveat, or question.
Lee et al. (2018) studied this question.